69-15-20
Kansas Administrative Regulation No. 69-15-20
69-15-20 Approved sterilization modes.
- Instruments used in the practice of permanent color, tattoo artist services,
or body piercing shall be sterilized, using one of the following methods:
- In a steam or chemical autoclave sterilizer, registered and listed with
the federal food and drug administration, and used, cleaned, and maintained
according to manufacturer's directions; or
- with single-use, prepackaged, sterilized equipment obtained from reputable
suppliers or manufacturers.
- Practitioners shall sterilize all piercing instruments that have or may
come in direct contact with a client's skin or be exposed to blood or body
fluids. Piercing needles shall not be reused. All piercing needles shall be
single use.
- All sterilizing devices shall be tested on a regular basis for functionality
and thorough sterilization by use of the following means:
- Chemical indicators that change color, to assure sufficient temperature
and proper functioning of equipment during the sterilization cycle; and
- a biological monitoring system using commercially prepared spores, to assure
that all microorganisms have been destroyed and sterilization has been achieved.
This testing shall be performed every three months for tattoo and body piercing
facilities.
- Chemical and biological indicator test results shall be made available at
the facility at all times for inspection by the board compliance officers.
- Practicing permanent color, tattoo, and body piercing facilities shall submit
to an inspection at least once every year, in order to determine compliance
with requirements of the statute and with sanitation and sterilization practices.
(Authorized by K.S.A. 1996 Supp. 74-2702a and implementing K.S.A. 1996 Supp.
65-1946, 65-1948, and 65-1949; effective Aug. 22, 1997.)